Basic setup
continuous pumping
START
STOP
The process control is not active, i. e. the controller is ready for
vacuum control, but control operation has not been started.
☞
The mode as from last operation is reactivated (after first
switching on ”vacuum control” (factory-set) is displayed).
☞
The actual pressure is displayed.
☞
Values as from last operation are reactivated.
☞
Press key START/STOP to start process control.
☞
If a coolant valve should be operated, the coolant valve has
to be preselected, see section ”Changing the basic setup”.
After switching on
Starting and stopping the process control
➨
Starting process control:
Press key START/STOP.
☞
If a coolant valve is preselected: Coolant valve opens im-
mediately, the symbol is displayed.
➨
Stopping process control:
Press key START/STOP.
☞
The pump is switched off.
☞
Attention:
Adopt suitable measures if there is a possible
pressure increase after switching off the pump, e. g. due to
low boiling solvents.
☞
If a coolant valve is preselected: The coolant valve continu-
ous to run for another 5 minutes to allow condensation of
vapours which remain in the system. Symbol ”cooling water”
is flashing.
Venting
(only possible if an external venting valve is connected and configurated)
VENT
Attention: If pressurized
inert gas is used install an
overpressure relief valve.
Max. permitted pressure
at controller: 2 bar
absolute. Ensure that high
pressure is compatible
with the mechanical
stability of the system!
Short venting
, e.g. in case of boiling retardation or if sudden
foaming occurs:
➨
Press key VENT shortly (< 2 sec.).
☞
Process control is stopped as if START/STOP was pressed.
Venting the system:
➨
Press key VENT continuously (> 2 sec.) until symbol ”vent-
ing valve” is flashing.
☞
Process control is stopped as if START/STOP was pressed.
☞
Ventilation until pressure does not change any more, i. e.
until atmospheric pressure is attained.
☞
Key START/STOP interrupts venting at any time.