To chose a pressure value for automatic switching off is
only possible if automatic switching off is activated (in ba-
sic setup vacuum control with automatic).
➨
Press key MODE simultaneously with
▼
(arrow down).
☞
When the key is pressed shortly, the current pressure for
automatic switching off is displayed for one second (”AE”,
i.e. automatic switching off, is preselected at the factory).
With a second tip within one second or continuous press-
ing:
➨
Select pressure for automatic switching off by using keys
▲
or
▼
(pressure selectable in a range from 1 to 1060 mbar
or ”AE”; ”AE”, i. e. automatic switching off, is active if a pres-
sure value < 1mbar is selected).
☞
The pressure is stored as new pressure for automatic switch-
ing off when the key is released.
Automatic switching off
MODE
30
mbar
Changing to basic setup continuous pumping
➨
Press key MODE simultaneously with
▲
(arrow up).
☞
Switching to continuous pumping is only possible, if vacuum
control is not activated (after pressing key START/STOP).
☞
In the basic setup continuous pumping the pumping speed
can be changed by presetting the motor speed.
☞
Pressing key MODE a second time: switching back to basic
setup vacuum control (only possible when pumping down).
MODE