How to change the pressure units
+
mbar
Torr
hPa
(mains switch)
Keep key p
▲
or p
▼
pressed during switching on.
☞
The pressure units are displayed, the pressure unit as from
last operation is flashing.
➨
Press key p
▲
or p
▼
to change the pressure unit. Press key
START/STOP when the controller displays the desired pres-
sure unit to finish the operation mode.
p
If the automatic mode has been preselected
”End”
is displayed.
☞
Switching between automatic switching off activated or not
activated is possible by pressing key
▲
or
▼
.
➨
Confirm by pressing key START/STOP.
☞
Automatic switching off as soon as the controller detects
that evaporation is completely finished. The pump is switched
off, the cooling water is running on, if a coolant valve is
preselected.
☞
Without automatic switching off, the pressure in the vacuum
system is adjusted after the evaporation is completed if nec-
essary until the ultimate vacuum of the system.
The controller switches to normal operation mode (process control not active).