Product installation
Installation environment
------------------------------------------
Product handling
1.
2.
3.
Protect the pharmacy refrigerator
when
moving it, same as shown as below
photo, please move it by handcart with cushion.
Remove all packing materials and bottom cushion, then move into
position
for placement.
After moving it to appropriate location, wait for 2 hours before power
ing
on.
---------------------------------------------------
The
pharmacy refrigerator is only for indoor use, with second pollution level and
second type of over-voltage
ambient temperature
tolerance
is within 16°C-32°C, and recommended
temperature is within 18°C-25°C. If necessary, air conditioning system should
be used.
Environmental humidity: lower than 80%Rh.
Avoid heavy dust.
Avoid mechanical sway or vibration.
The altitude of working position of
refrigerator is lower than 2000m
.
Input voltage: between 187v~242v
Ingress protection
:
IPX0.
The pharmacy
i
environments than the one aforesad, the pharmacy refrigerator can't operate. normally. Please improve the
environment before use.
I t is prohibited to install the
pharmacy refrigerator in an open environment.
.
When the pharmacy refrigerator is wet by rain, it may cause electric leakage
or electric shock.