Precautions for use
Before putting
items
into the device, please make sure that the
internal
temperature in the
cabinet
has reached the set temperature, and then put
the
items
into the pharmacy refrigerator .i n batches.
The volume of the
articles in the
box should not exceed 1/4 of its volume at a time to prevent
the temperature
from rising too much.
The temperature display value of the device is to show the temperature at
the temperature sensor in the storage room. There is a certain difference
between the displayed temperature and the actual temperature at the center
of the device when the device just starts. As the device enters a stable state,
the displayed temperature will gradually reach the actual temperature.
The device box body is provided with a detection hole so as to lead out
the test line in the box during the test. After the test line is led out, it is
necessary to plug the test hole with insulating material again, otherwise
the temperature in the box may not reach the set temperature value, and
condensation will appear around the outside of the through hole.
Please use diluted neutral cleaner to clean the device. Do not use brushes,
acid, gasoline, soap powder, polishing agent or hot water to clean the
device, otherwise the painted surface and plastic rubber parts may be
damaged. Attention: Do not use volatile solvents such as gasoline to clean
plastic and rubber parts.
When the device is not in use for a long time, cut off the power supply and
turn off the battery.
Each time you access the items, please try to reduce the time to open the
door to avoid large fluctuations in temperature and humidity inside the box.
After the door is opened, the temperature inside the box will rise sharply in
a short time. This is normal, and the situation will be restored within 1 hour
after the door is closed.
Under high temperature and high humidity environment, glass door may
condense, which is normal phenomenon and does not affect the storage
temperature in the box.