Diagnostics and servicing
page 49
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
F265
Measurement func-
tion disturbed
Program sequence of the meas-
uring function disturbed
Device restarts automatically
Byte 4, Bit 3 of
Byte 0 … 5
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
C700
Simulation active
A simulation is active
Finish simulation
Wait for the automatic end after
60 mins.
"Simulation Active"
in "Standardized
Status 0"
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
S600
Impermissible elec-
tronics temperature
Temperature of the electronics in
the non-specified range
Check ambient temperature
Insulate electronics
Byte 23, Bit 4 of
Byte 14 … 24
S601
Overfilling
Danger of vessel overfilling
Make sure that there is no fur-
ther filling
Check level in the vessel
Byte 23, Bit 5 of
Byte 14 … 24
S603
Impermissible op-
erating voltage
Terminal voltage too small
Check terminal voltage, increase
operating voltage
Byte 23, Bit 6 of
Byte 14 … 24
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
M500
Error in the delivery
status
The data could not be restored
during the reset to delivery status
Repeat reset
Load XML file with sensor data
into the sensor
Bit 0 of
Byte 14 … 24
M501
Error in the non-
active linearization
table
Hardware error EEPROM
Send instrument for repair
Bit 1 of
Byte 14 … 24
M507
Error in the instru-
ment settings
Error during setup
Error when carrying out a reset
False signal suppression faulty
Carry out reset and repeat setup Bit 7 of
Byte 14 … 24
M508
No executable
Bluetooth software
Checksum error in Bluetooth
software
Carry out software update
Bit 8 of
Byte 14 … 24
Function check
Out of specification
Maintenance