NivoRadar
®
Continuous level measurement
Series NR 7200
Technical information / Instruction manual
Set up with the integrated display and adjustment unit
page 34
Releasing the sensor adjustment is also possible in any menu item
by entering the device code.
Note:
The factory set device code is " 000000". If this is changed by
the user and is no longer available, access is only possible via the
emergency device unlock code on the information sheet " Emer-
gency unlock codes" also supplied.
Caution:
With protected parameter adjustment, adjustment via the adjust-
ment tool and other systems is also blocked.
During a reset, parameter settings made by the user are reset to
the values of the basic setting or the delivery status (see chapter "
Menu overview")
1)
.
Caution:
For the duration of the reset, the set trouble signal is output via the
current output. Within the context of the asset management func-
tion, the message " Maintenance" is output.
The following reset functions are available:
Basic settings:
Reset the parameter settings to the default values of the respective
device. You will find the values in chapter " Menu overview".
Note:
Order-related settings are not taken over into the current parameters
after this reset. The currently set menu language is not reset.
Delivery status:
Resetting of the parameter adjustments to delivery status
Country specific settings for the radar signals are determined via
the operating mode.
Reset
Mode
1) Language and Bluetooth access code are not reset.