4. INITIAL SETUP
4.1 PLACING THE SCALE
In order to obtain an accurate weighing result, the scale
must be placed on a strong and level surface.
If required, adjust the adjustable feet underneath
scale
to obtain a level position. Refer to the bubble indicator
for leveling condition.
4.2 INTERNAL SETTING
4.2.1 Display Segment Check
User can enter this function to check to see if the display
and backlight (if purchased) are functioning properly.
a. Scale is off
b. Press and hold
SAMPLE SET
, then press
ON/RE-ZERO
c. Scale displays
F.1
d. Press
SAMPLE SET
once
e. Scale displays
F.2
f. Press
MODE
and all segments will light up to allow
user to check the condition of display
g. Press
SAMPLE SET
for next function setup or
ON/RE-ZERO
to restart the scale
4.2.2 Select Auto Power Off Setting
The scale is equipped with
Auto Power Off Function
. The
scale will power off automatically if it is not used
for 4 minutes. Follow the below steps to enable/disable
Auto Power Off Function.
a. Scale is off
b. Press and hold
SAMPLE SET
, then press
ON/RE-ZERO
c. Scale displays
F.1
d. Press
SAMPLE SET
three times
e. Scale displays
F.4
f. Press
MODE
to shift between
0._OFF
and
4._OFF
-To enable
Auto Power Off Function
, press
SAMPLE
SET
when
4._OFF
appears
-To disable
Auto Power Off Function
, press
SAMPLE
SET
when
0._OFF
appears
g. Press
SAMPLE SET
for next function setup or
ON/RE-ZERO
to restart the scale
4.2.3 Select RS-232 Baud Rate & Protocol
This scale is equipped with 3 RS232C output baud rate
and 2 transmission protocol settings for user to select.
a. Scale is off
b. Press and hold
SAMPLE SET
, then press
ON/RE-ZERO
c. Scale displays
F.1