17 

Transilluminator Will Not Turn On 

Be sure the darkroom cabinet’s door is completely closed. There is a UV exposure safety cut-off switch 

that turns the transilluminator off when the darkroom cabinet’s door is opened. 

Be sure the darkroom cabinet’s main power switch is lit. If not, refer to “No power to darkroom cabinet.” 

The transilluminator itself has a power switch; make sure that the transilluminator switch is in the On 
position. 

Be sure the transilluminator’s power jumper cord is securely connected at both the ends of the cable 

There is a 10-minute timer that turns off the transilluminator automatically after 10-minutes of use. The timer 
switch located at the front of the darkroom can be reset (which restarts the 10-minute count).  

 

Error Messages Pop Up on the Screen 
 

Messages appear on the screen that might be related to the TS software interface, or Microsoft Windows. If 
the message is related to Microsoft Windows, such as a reminder to activate or update your copy of 
Windows, please contact your system administrator for assistance. 

If an error message appears repetitively, and your system administrator does not recognize it as a Microsoft 
Windows error, contact UVP Technical Support for further assistance. 

 
Technical Support 

 

UVP offers technical support on all of its products.  If you have any questions about the product’s use, 
operation or repair, please contact our offices at the locations below. Or go to UVP’s web site and click the 
Tech Support > BioImaging Systems.  

 

If you are in North America, South 

America, East Asia or Australia: 

If you are in Europe, Africa, the 

Middle East of Western Asia: 

Call (800) 452-6788 or (909) 946-

3197

, and ask for 

Customer Service

 

during regular business days, between 
7:00 am and 5:00 pm, PST. 

Call +44(0) 1223-42002

, and ask for 

Customer Service

 during regular 

business days between 9:00 am and 
5:30 pm. 

E-mail

 your message to: 

[email protected] 

E-mail

 your message to: [email protected] 

Fax

 Customer Service, and send it to 

(909) 946-3597

 

Fax 

Customer Service, and send it to: 

+44(0) 1223-420561

 

Write to:

 

UVP, LLC

 2066 W. 11

th

 

Street, Upland, CA 91786 USA 

Write to

Ultra-Violet Products Ltd

  

Unit 1, Trinity Hall Farm Estate, Nuffield 

Road, Cambridge CB4 1TG UK 

 
 

Summary of Contents for BioDoc-It

Page 1: ...Products Ltd 2066 W 11th Street Unit 1 Trinity Hall Farm Estate Upland CA 91786 Nuffield Road Cambridge CB4 1TG UK Phone 800 452 6788 909 946 3197 Phone 44 0 1223 420022 Fax 909 946 3597 Fax 44 0 1223...

Page 2: ...transilluminator in the lab Configurations include a compact hood or camera stand The compact hood fits on top of UVP s High Performance and Benchtop Transilluminators The VisiDoc It System is simila...

Page 3: ...efer to the packing slip and pictured components for parts included with the system VisiDoc It with Hood Transilluminator check the packing slip to determine if the system includes a transilluminator...

Page 4: ...ransilluminator with UV Blocking Cover check the packing slip to determine if the system includes a transilluminator FluorCam 210 Camera with Lens Kit EtBr Filter is Mounted under the Lens 8 LCD Touch...

Page 5: ...front 4 in the back USB storage device capacity 1GB minimum Software TS software Ethidium Bromide Filter EtBr The Ethidium Bromide EtBr Filter 50mm sq UV blocking band pass interference filter blocks...

Page 6: ...hs and high low intensities or the high end FirstLight transilluminator Note For UV protection there is a 10 minute timer installed in the BioDoc It system The timer shuts off the transilluminator aft...

Page 7: ...aviolet transilluminator The converter plate is specially coated to convert the 302nm UV output to white light rather than using a separate white light box The Visi Blue Converter Plate converts UV to...

Page 8: ...m is turned OFF during cleaning Keep the equipment away from motors or other large magnetic equipment apparatus Installing the Filter Systems include a 50mm sq EtBr filter The filter is pre installed...

Page 9: ...nly fit one way Using the four thumb screws provided secure the bracket to the base Slide the camera and lens assembly into the camera bracket Slide the lens through the center hole in the gasket The...

Page 10: ...her end into a UBS port on back of the darkroom Connect the jumper cord from the darkroom to the transilluminator Connect the main power cord from the back of the darkroom to a surge protector power o...

Page 11: ...diopter will only fit one way Using the four thumb screws provided secure the bracket to the base Slide the camera and lens assembly into the camera bracket Slide the lens through the center hole in...

Page 12: ...nect the power cord from the transilluminator to a surge protector power outlet Connect the main power cord from the back of the processor to a surge protector power outlet Printer Set up The thermal...

Page 13: ...er Using the System Powering Up the System In the BioDoc It Power up the system by pushing once on the MAIN POWER button located on the right side of the unit All internal components will now have pow...

Page 14: ...desired angle Turn each knob clockwise to tighten Operating the TS Software The system will go through the boot up process When ready the LCD will display the TS software screen similar to the one be...

Page 15: ...any exposure to UV Using the Gel Viewing Window in the BioDoc It The Gel Viewer has a pressure sensitive clasp Press firmly to open the viewer The glass is UV blocking while providing a clear view to...

Page 16: ...01 Fluorescent Standard Step Tablet 33 0014 02 Doc It LS Analysis Software 97 0185 02 Spectacles UV Blocking UVC 303 98 0002 01 Goggles UV Blocking UVC 503 98 0002 02 Faceshield UV Blocking UVC 803 9...

Page 17: ...r copy of Windows please contact your system administrator for assistance If an error message appears repetitively and your system administrator does not recognize it as a Microsoft Windows error cont...

Page 18: ...rument or part thereof that has been subject to accident negligence alteration abuse or misuse by the end user Moreover UVP makes no warranties whatsoever with respect to parts not supplied by UVP or...

Reviews: