Uvitron International INTELLI-RAY 600 Instruction Manual, Rev. T
Page 18 of 36
SYSTEM OPERATION (CONTINUED)
6
If a value of 0 is entered for exposure time, then Manual Exposure will begin. This will result in
the shutter remaining open indefinitely, until a subsequent press of the ENTR key occurs to end
the exposure. During Manual Exposure, the Run Screen timer will count up, indicating the total
time that has elapsed.
NOTE:
Any time the shutter is open, it can be closed immediately (exposure turned
OFF) by pressing the ENTR key.
If the MENU key is pressed twice while exposure is in progress, the Exposure Time screen will
be displayed, allowing the time setting to be changed without closing the shutter. If in either
Timed or Manual Exposure mode, the amount of time already elapsed will be subtracted from
the newly entered value, so that the total time of exposure will be equal to the new value
entered. If the time elapsed is greater or equal to the new value entered, the exposure will
terminate, and the shutter will close.
M6
Lamp/Oven Temperatures
This screen will display the internal temperature of the IntelliRay lamp head. If the IntelliRay is
being used with a Rayven Oven, then the Rayven internal temperature will also be displayed.
Note that in order for the oven temperature to be displayed, the DSUB interface cable must be
connected between the rear connectors of the IntelliRay and Rayven, and the door interlock
must be enabled in the Setup menu (Interlock ON/OFF item S7). If the interlock is not enabled
or if the Oven Over-temperature alarm is disabled by a setting of less or equal to 0°C (32°F) in
the Setup menu, then the word OFF will be displayed for oven temperature. The units for the
temperature display can be selected as degrees F or C in the Temp Unit Setup menu item S8.
M7
Alarm Log (Appears Only When Alarm(s) Active)
When the IntelliRay control system detects an alarm condition has occurred, an audible beeping
will sound and a pop-
up “ALARM PRESENT, ENTR to View” message will be sent to the LCD
display. To view & acknowledge the active alarm and silence
its’ beeping, press the ENTR key.
As long as this or any other previously detected alarm condition persists, an extra Alarm Log
menu item M7 will be present at the end of the main menu that will identify the existing
problem(s) with appropriate messages on the LCD display. See System Alarms in section 9.2
for more information.
M8
Shutter Open (Run Screen)
When timed or manual exposure has been started from the Exposure Time screen, then the run
screen is automatically displayed, showing the following system parameters:
Intn = Current lamp intensity
Timer = Current value of Exposure Timer
Hours = Age of arc lamp in hours (not displayed when exposure timer set for Hours)