UVITEC – Cambridge
209/225
Display – Image display pre-defined parameters
The Display folder contains the image display pre-defined set-up:
Select display mode
The Select display mode options allow you to select between a standard display, an enhance display
for the faint band and an enhance display for the bright band.
Standard display is optimized for all type of image.
Enhance display for faint band ensure the faint bands are more visible, but more prominent bands
will look very strong. This is well adapted for chemiluminescence sample.
Enhance display for bright band ensure the bright bands are more visible compared to their
background. This is well adapted for fluorescence sample.
Set the gamma value
Gamma adjustment corrects an image by creating a new version of the original. To create the new
image, the Gamma Adjust function reassigns the grey values of each pixel in the image according to
the curve in the following graph: