General sales and delivery conditions of UV-Technik Speziallampen GmbH are valid
.
Provision of the manual is mandatory
.
page 12 of 17
subject to change without notice
operating manual EVG UVT 200-480W
Monitoring circuitry
mains voltage monitoring
cut off in case of persisting under- or overvoltage
temperature monitoring
cut off in case of persisting over temperature
lamp presence check
filament check
prevention of ignition if lamp is not connected or
if a wrong filament is indentified
short circuit of lamp wires
immediate cut off
lamp failure
(see possible causes at status indication)
cut off
Status indication - generally
operation indicator
LED green
normal operation
trouble indicator
LED red
failure by cut off
potential free contact (failure contact)
relay / changer
(terminal 6-8, 9-11)
relay on
(6-7 closed, 9-10 open)
relay off
(7-8 open, 10-11 closed)
normal operation
no mains / no lamp ignition / failure
Status output
external LED status indication
(RJ45)
supplied via internal 12 V DC
(output short-circuit protected)
galvanically isolated connection of 2 external LEDs per lamp channel
series resistor must be calculated according to the LED current, minimal 600 Ohm
1: +12V
2: +12V
3: GND LED red channel 1
4: GND LED red channel 2
5: GND LED green channel 1
6: GND LED green channel 2
optional possibilities
(RJ45, terminal 20)
RS485 interface
Installation instructions
designed for
installation in electrical cabinets
IP protection class
IP20
size of housing (w x d x h)
borehole spacing 240 mm
250 x 105 x 60 mm
(2x 100-200W)
250 x 150 x 60 mm
(2x 150-320W)
mounting position
vertical
(mains supply connector downside for optimal heat dissipation)
ambient temperature
t
a
= 0...40°C
relative humidity
max. 80% non condensing
temperature at tc - point
t
c
= 50°C max. at housing
(forced cooling necessary if exceeded)
Wire length / cabling
allowed wire length
to 5 m
(longer cable possible, depending from cable capacity and laying)
allowed cable capacity
max. 750 pF/m
shielded wire allowed
yes
(Attention: Connect shield only at one side, avoid ground loops!)
Connectors
clamp
connector for
allowed stranded wire gauge with ferrule
allowed rigid wire gauge
1, 2
remote input
Ø: 0.25...1.0 mm² / 30-17 AWG
0.2-1.5 mm² / 32-16AWG
3..5
mains
Ø: 0.75...1.5 mm² / 18-16 AWG
0.75-2.5 mm² / 18-11AWG
6..8, 9..11
relay contacts
Ø: 0.25...1.0 mm² / 30-17 AWG
0.2-1.5 mm² / 32-16AWG
12..15, 16..19
lamps
Ø: 0.75...1.5 mm² / 18-16 AWG
0.75-2.5 mm² / 18-11AWG
20
(RJ45)
output (LED/485)*
optionally: accessory cable 1 m, with RJ45 plug
* only active in ballast types with flexible adjustment
Compliance with standards
Safety
EN 61347-2-3
EMC limits for harmonic current emissions
EN 61000-3-2
EMC radio interference suppression
EN 55011 class A1
EMC immunity
EN 61547
The stated EMC standards apply only to single ballast and for cable length < 3 m. Since the ballast is a part of an installation it
needs in most cases an extra EMC approval of the whole device and necessarily additional measures to reduce disturbances.
RJ45 connector