USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-24
6.
Place the
CENTRAL
tile face down on the color head and press
the
Start
button
(on the Mic cabinet)
.
The software will indicate when calibration is complete and a PASS
condition has been achieved.
ATTENTION:
Make sure the tiles are observed in the order requested
on the screen. During this procedure the tile being
measured is compared to the standard value stored for
that tile. If the values are different, the system adjusts
the constants, which could cause measurements to be
skewed.
Note:
Always use the same set of tiles for daily calibration that you
used for original color calibration.
DO NOT MIX TILE SETS
.
The software automatically compares the color of the tiles to the
standard values recorded for those tiles. If necessary, it adjusts system
constants.
When calibration is complete and within tolerance, you can view or
print the
Report
.