USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 6
(3/08) System
Testing
-
Page 6-19
Discrepant data includes any combination of side limit or lot limit results.
Side limits are indicated by a
RED
background. Lot Limits are indicated
by a
BLUE
background.
Note:
Discrepant means the bale has had one (1) Lot Limit or Side
Limit result.
If the Length/Strength, Color, or Trash test results indicate that the
measurements from the two sides of the bale are not within the tolerance
limits, an error message is displayed beside the measured value for the
field out of tolerance. A window will launch and ask if you want to
Accept
or
Retest
.
After retesting, the system automatically accepts all additional testing.
If, after the cotton sample is retested, the values are still outside the
tolerance limits, this is an indication that the cotton sample came from a
"between-sides"
bale.
The following conditions indicate that retesting may be necessary:
1.
An inadequate length/strength amount (high/low cotton sample
amount) is measured. In this case, the amount of cotton in the
offending sampler drum may be adjusted or the cotton sample
repositioned. Pressing the
L/S Start
button will measure two more
combs. Only the data from the sampler drum(s) with the incorrect
amount will be utilized.
2.
A lot limit is violated. If the lot limit is associated with a module,
only that module must be retested. If a lot limit is associated with
several modules, such as SCI or maturity, then all parameters must
be measured again. However, the Bale ID is retained. The
sequence can be explained as follows:
o
If a lot limit occurs with length/strength, click
Retest
and press
the
Start
button on the L/S cabinet. Only data from the
sample that violated the lot limit will be considered.
o
If a lot limit occurs with color/trash, click
Retest
and press the
C/T Start
button on the micronaire cabinet. Only data from
the cotton sample that violated the lot limit will be considered.