Length/Strength Control Board
Error
Code
Error Code
Error Code
7
Front Jaw NOT Open Before Test.
8
Rear Jaw NOT Open Before Test.
9
Front Jaw NOT Closed For Test.
10
Rear Jaw NOT Closed For Test.
1. Check to insure that there is
an adequate air supply.
2. Go to L/S CTRL Board
Diagnostics and
OPEN/CLOSE the front
and/or rear jaws using the
buttons in the Digital Outputs
column. Notice if sensors
turn ON/OFF in the Digital
Inputs column.
11
Moisture Measurement Result NOT
Received.
Check for loose cable at the
comb transport and at the L/S
Control Board.
12
Front Jaw NOT Open After Test.
13
Rear Jaw NOT Open After Test.
See 7-10 above.
14
Track Motor State Sensor Error.
1. Restart the HVI Program.
2. Go to L/S CTRL Board
Diagnostics and turn the
Track Motor ON/OFF using
the buttons in the Digital
Inputs column. Notice if
sensors turn ON/OFF
(above). If sensors
unresponsive, may be a
cable, sensor or control
board problem.
See Diagnostics Section for
details
15
Track Motor Power OFF Error.
1. Restart the HVI Program.
2. Go to Track Motor
Diagnostics and turn the
Track Motor OFF and check
the status byte. If not
responsive may be a
problem with the motor,
cable or sensor.