30
1000033244-00
4. Instructions for use
Model UVD-C254
UVD-C365
UVD-C405
Sensitivity wavelength
region (nm)
220 ~ 310
310 ~ 390
320 ~ 470
Wavelength for
calibrating absolute
value (nm)
254 365 405
Photodetector diameter
(mm)
φ
10
φ
10
φ
10
Non-linearity
Within
±
1%
Within
±
1%
Within
±
1%
Temperature measured
range (
°
C)
0 ~ 50
0 ~ 50
0 ~ 50
H range
0 ~ 9999
0 ~ 9999
0.0 ~ 999.9
M range
0.0 ~ 999.9
0.0 ~ 999.9
0.00 ~ 99.99
Irradiance
measured
ranges
(mW/cm²)
L range
0.00 ~ 99.99
0.00 ~ 99.99
0.000 ~ 9.999
H range
0 ~ 99999
0 ~ 99999
0.0 ~ 9999.9
M range
0.0 ~ 9999.9
0.0 ~ 9999.9
0.00 ~ 999.99
Accumulated
light amount
measurement
range
(mJ/cm²)
L range
0.00 ~ 999.99
0.00 ~ 999.99
0.000 ~ 99.999
Spectroscopic sensitivity
characteristics
Model UVD-S254 UVD-S365
UVD-S405
Sensitivity wavelength
region (nm)
220 ~ 310
310 ~ 390
320 ~ 470
Wavelength for
calibrating absolute
value (nm)
254 365 405
Photodetector diameter
(mm)
φ
3
φ
1
φ
1
Non-linearity
Within
±
1%
Within
±
1%
Within
±
1%
Temperature measured
range (
°
C)
0 ~ 50
0 ~ 50
0 ~ 50
H range
0 ~ 9999
0 ~ 9999
0.0 ~ 999.9
M range
0.0 ~ 999.9
0.0 ~ 999.9
0.00 ~ 99.99
Irradiance
measured
ranges
(mW/cm²)
L range
0.00 ~ 99.99
0.00 ~ 99.99
0.000 ~ 9.999
H range
0 ~ 99999
0 ~ 99999
0.0 ~ 9999.9
M range
0.0 ~ 9999.9
0.0 ~ 9999.9
0.00 ~ 999.99
Accumulated
light amount
measurement
range
(mJ/cm²)
L range
0.00 ~ 999.99
0.00 ~ 999.99
0.000 ~ 99.999
Spectroscopic sensitivity
characteristics
Integrated-type photodetector Specifications
Separate-type photodetector Specifications
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
200
250
300
350
Wavelength(nm)
Relative spectroscopic sensitivity
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
300
350
400
Wavelength(nm)
Relative spectroscopic sensitivity
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
300
350
400
500
450
Wavelength(nm)
Relative spectroscopic sensitivity
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
200
250
300
350
Wavelength(nm)
Relative spectroscopic sensitivity
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
300
350
400
Wavelength(nm)
Relative spectroscopic sensitivity
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
300
350
400
500
450
Wavelength(nm)
Relative spectroscopic sensitivity
8. Specifications