19
1000033244-00
4. Instructions for use
(3) Press the
switch once.
-
is indicated in display monitor.
- All indicators are turned off.
(4) When the
switch is pushed, the irradiance distribution is output on the
analog voltage signal.
-
is blinked on the display monitor.
- The irradiance distribution is output using the 0
−
1 V of the analog voltage
signal. When the irradiance value is 0 the analog signal is 0 V. When the
irradiance value is at full scale, the analog signal outputs 1 V.
- The temperature distribution is output in the range of the analog voltage sig-
nal of 0
−
1.75 V. When the temperature is 0
°
C, the analog signal output is
0 V. When the temperature is 200
°
C, the analog signal is 1 V.
- Reference "4-3-1. Analog output" for irradiance and temperatures when ana-
log voltage is 1 V.
(5) When the output of the analog voltage signal is complete, the peak irradiance
value or peak temperature is indicated on the display monitor.
- The irradiance distribution is output for 2 minutes for either sampling rate
(16 or 32 samples/minute).
The distribution output is coarser for a sample rate of 16 samples/second than
it is for 32 samples/second.
(6) If you would like to know the accumulated exposure amount for the irradiance
distribution, push the
switch 3 times.
- Accumulated measurement mode is entered.
- The accumulated exposure amount value is indicated on the display monitor.
(7) If the
switch is pressed during output of the analog voltage signal the
output is forcefully stopped.
- The peak irradiance or peak temperature value is indicated on the display
monitor.