14
1000033244-00
4. Instructions for use
4-2-2. Turning unit off
(1) Push the
switch for roughly 1 second.
- Numbers on the display monitor disappear.
- The power is turned off.
4-2-3. Real time irradiance (temperature) measurement
(1) When the UIT-250 is turned on, it enters H range real time irradiance measure-
ment mode.
- The real time irradiance indicator is blinking.
- The average value for 0.5 seconds is indicated on the display monitor.
- When a temperature sensor is connected, the thermocouple temperature is
indicated on the display panel.
(2) Push the
switch to change the measurement range.
- The measurement range changes from H range
→
M range
→
L range
→
H range each time the switch is pushed.
- Reference table 1 for display of each measurement range and for full scale
values.
Note, the measurement range is not selected when a temperature sensor is
connected.
Table 1 Indication ranges of irradiance values and temperature
Model
UVD-C254
UVD-S254
(mW/cm
2
)
UVD-C365
UVD-S365
(mW/cm
2
)
UVD-C405
UVD-S405
(mW/cm
2
)
UVD-TK
(
°
C)
H range
0 ~ 9999
0 ~ 9999
0.0 ~ 999.9
M range
0.0 ~ 999.9
0.0 ~ 999.9
0.00 ~ 99.99
Indication
limits
L range
0.00 ~ 99.99
0.00 ~ 99.99
0.000 ~ 9.999
0.0~350.0
(3) When the
switch is pushed, the real time irradiance value is held.
- The real time irradiance indicator light turns on.
The hold can be canceled by pushing the
switch a second time.
- When in analog output, output is continuous whether the
switch is pushed
or not.