Document: MAN00238A
Page 26 of 28
R:\Documents\MAN Manuals\MAN00238A BRC-15-1111R-STD-J37.doc
Troubleshooting
A. Equipment
Resistor Load:
13.5k ohms, 16.7kW
Current Sensor:
Tektronix Current Probe #P6021
Voltage Sensor:
Voltage Divider, 20kV minimum
Ratio: 500:1 or greater
Frequency Response: 100kHz minimum
Resistance: 10 megohms typical
Scope Probes: 10:1 and 100:1
Test Equipment:
Tektronix Oscilloscope, digital storage recommended
Fluke
Multimeter
B. Initial
Turn-on
1.0
The ‘Power ON’ LED on the front panel board should be illuminated. If the front
panel shows no lights, check the following:
1.
Check for power on the input power line (208VAC, 3-phase).
2.
Check the fuses on Regulator board 9120010 (see section 8.8 for fuse
details).
2.0
It is normal for the front panel ‘Fault’ LED to blink momentarily when the circuit
breaker is turned on. The ‘Fault’ LED should immediately go out, however if the
light does not extinguish, check if any fault LEDs are illuminated on the rear
panel. If one or more LED is illuminated, verify if the fault condition(s) is
present and correct if possible.
3.0
High Voltage Ready Status
The green ‘Ready’ indicator in the ‘HV ON’ pushbutton should be illuminated
when conditions are clear for HV on. If faults are present, this indicator will not
illuminate.