7
2
CONNECTION
7
CONNECTION
Chapter
2
2-2.
Connection of the strain gauge sensor
Connect a strain gauge sensor.
The excitation voltage is 2.5V or 5V. (designated when ordered)
The maximum output current is 30mA.
■
4-wire strain gauge sensor
■
6-wire strain gauge sensor
For connecting a 6-wire strain gauge sensor, short-cEXC and +S, and -EXC and -S.
Signal input/output
1 2 3 4 5
+EXC
-SIG
-EXC +SIG
2
3
4
5
Strain gauge sensor
SHIELD
1
terminal block
With
mark: The excitation voltage is 5V.
With no mark:
The excitation voltage is 2.5V.
BV5
Point
(+EXC)
(-EXC)
(+SIG)
(-SIG)
(SHIELD)
+IN
-OUT
-IN
+OUT
2
5
F331
1
3
4
(+EXC)
(-EXC)
(+SIG)
(-SIG)
(SHIELD)
+IN
-OUT
-IN
+OUT
2
5
F331
1
3
4
Summary of Contents for F331
Page 1: ...OPERATION MANUAL 26MAY2014REV 1 02 DIGITAL INDICATOR F331 ...
Page 79: ...70 8 SPECIFICATIONS 70 SPECIFICATIONS Chapter 8 8 2 Dimensions Unit mm ...
Page 82: ...73 8 SPECIFICATIONS 73 SPECIFICATIONS Chapter 8 When the RS 485 option is equipped Unit mm ...
Page 83: ...74 8 SPECIFICATIONS 74 SPECIFICATIONS Chapter 8 When the RS 232C option is equipped Unit mm ...