3
1
OUTLINE
3
OUTLINE
Chapter
1
■
Rear panel
①
Protective ground
This is a protective ground terminal block. Be sure to ground the protective ground terminal to
prevent electric shocks and failures by static electricity. (The frame and protective ground terminal
are conducted.) Do not use other screws than that attached to the main body (M4×8 binding-head
machine screw with a toothed washer).
②
Signal input/output terminal block
This terminal block is used for input/output of control signals and input of strain gauge sensor
signals.
- Terminal block assignment
1 to 5:
Terminal for connecting a strain gauge sensor.
1
SHIELD
2
+EXC
3
-SIG
4
-EXC
5
+SIG
6
・
7:
Terminals for inputting Hold/Judge signals.
6
COM
7
IN1 (Hold/Judge input)
6
・
8:
Terminals for inputting digital zero signals.
6
COM
8
IN2 (DZ input)
②
Signal input/output terminal block
①
Protective ground
③
Option slot