16.SPECIFICATIONS
109
16. SPECIFICATIONS
16-1. Analog Section
Sensor excitationvoltage
10V or 2.5V DC±10%
(The factory-shipped initial value is 2.5V.)
Output current within 30mA
Signal input range
-3.0 to +3.0mV/V
Equivalent input calibration range
+0.5 to +3.0mV/V
Equivalent input calibration error
Within 0.1
%
FS
(
0.5mV/V input
)
Zero adjustment range
0 to
±
2.0mV/V
Gain adjustment range
+0.5 to +3.0mV/V
Analog input signal sensitivity
1
μ
V/count
(1/10000 can be assured with 1mV/V
or more input.)
Accuracy
Non-linearity Within 0.02%FS
(
of 3mV/V input
)
Zero drift
Within 0.5
μ
V/
℃
Gain drift
Within 25ppm/
℃
A/D converter
2000 times /sec. Resolution:24 bits
(
binary
)
Analog filter
3, 10, 30, 100, 300, 1kHz
Peak hold function
Digital peak hold at 2000 times/sec.