6.SETTING OF FUNCTIONS
52
6-7. Analog Filter
This is a lowpass filter filtering the strain gage sensors input signal and cutout
the noise element.
Lowpass filter cutout freguency is selectable in the 3/10/30/100/300/1k Hz.
1)
Select setting mode 2.
2)
Select analog filter.
Use
and
keys to set the analog filter,
then use
key to validate the setting.
To return to the indicated value display, press
key.
FNC
+
-
ZERO
Press twice.
CAL
HOLD
GAIN
CAL
HOLD
FNC
+
-
Setting of Analog Filter
5 : 1kHz
4 : 300Hz
3 : 100Hz
2 : 30Hz
Analog Filter
1 : 10Hz
0 : 3Hz