32
5.7
Voltage Mode:
⚫
The instrument also has a voltmeter that measures AC/DC voltage from 20 V to 600 V.
This voltmeter is useful as it shows the presence of voltage when it is connected to a DUT
which has induced voltage.
⚫
To enter the voltage mode press the button on the home screen of the meter. The
following screen will be displayed.
⚫
Additionally the meter also displays the voltage at the test leads/DUT in other
modes(IR,DAR,PI,STEP and RAMP) before the start of the test as shown in
figure(figure
on next page).
⚫
Before the start of the test if voltage above 50V is present at the terminals the speaker
will make beep sound to indicate the presence of voltage in these modes(IR,DAR,PI,STEP
and RAMP).
VOLTAGE AT
LEADS/DUT