Unihedron
SQM-LU Users manual
4.2.8 Manually set light calibration temperature
Calibration is done at the factory, however, in the case where calibration values must be restored or set
to something else, this command allows a new calibration value to be placed into the SQM-LU.
Executing the command “
zcal6########.##x
” manually sets the light calibration temperature to
the value specified in “
########.##
”. The units are ºC.
Note: The SQM-LU records the temperature in a raw value with different resolution, so the reply back
may not be exactly the same as the value sent.
The format of the response is:
Column Example value
Description
0
z
Calibration response is being returned.
2
6
Manual Set Light Calibration Offset
4-9
019.0C
Value that was set into EEPROM
10-11
Carriage return (0x0d), Line feed (0x0a).
Table 9: Response for manually setting of light calibration temperature
An example is:
z,6,019.0C
0123456789
10
1
Revision 1.1
Page 15
Last modified: Jun 18, 2011