![Unico S1205 SERIES User Manual Download Page 21](http://html1.mh-extra.com/html/unico/s1205-series/s1205-series_user-manual_3734673021.webp)
21
Reset Defaults:
Reserved for future application;
Memory Manager:
It has three functions as shown.
A.
Load Saved data
Select
Load Saved data
. The saved data will all be listed.
With * next to the saved data interested (COBALT in this case), press
ENTER
, you
can print the data (if printer installed) and display it on screen, or you can send the data
to computer .
B.
Clear User Tests
Select
Clear User Tests
, all established and saved tests/curves will be deleted.
Warning:
All saved test curves will be deleted PERMANENTLY!!!
C.
Clear Data Memory
Select
Clear Data Memory
; all saved tests data will be deleted.
Warning:
All saved data will be deleted PERMANENTLY!!!
Wavelength Correction
: If wavelength is found to be not accurate. Use this function to make
the proper wavelength correction. Select
WL Correction
.