15
9.
Save the Curve defined: with the * at “Save Curve” press ENTER to save the define
curve. Then the * moves to “Dis & Print Curve” automatically.
10.
Display and Print-out Defined Test (Curve parameter): with the * at “Dis & Print Curve”
Press
ENTER
to display and print (if printer is connected) the defined test parameter.
The data and parameters include:
Defined test sequence number in All Test List. (No.=002 in this case);
The wavelength used for the defined test (581nm in this case);
The defined test name (BROMINE LR in this case);
Fitness method for the defined test ( “1” represents “1st order linear”, “0” represents “ 1st
order linear through zero”);
Intercept value (1.058 in thise case);
Slope number (2.588 in this case) and
Concentration unit (ppm in this case).
Note
: N and M are not part of parameter and have no real meaning and effect to the test.
The establishment of the defined test is complete and saved in the instruments. Now you
may make final check on all the parameters. Press EXIT to return to
All Test List
menu.
The “To be defined 002” has been replaced with “BROMINRE LR”.
Define a Test by Known Standards
If you have standards with known concentration values you can establish and define test curve
by following the instructions in this section.
1.
At Quantitative menu scroll * to the left of desired location for saving the test method.
Press
ENTER
to confirm.