18
If the “No. of Replics” is “3”, then place the second solution of STD No.1 in the
cuvette holder. Pressing
ENTER
will measure the second solution of STD N0.1
and the second measurement value will be displayed at STD No.1= 0.18775A2.
Repeat the same procedures to measure the third solution of STD No.1 and screen
will display STD. No.1=0.1876A3. This finishes the measurement of the three
solutions of standard No. 1.
Press
ENTER
again to average the measured absorbance values and the screen
then displays its averaged value: 0.1876A (0.0.0.1876A3)/3.
This averaged absorbance value will be used for calculation.
8.5
Place standard solution No.2 into the cuvette holder in the optical path. Repeat step
8.3 to 8.4 to measure standard No.2. Repeat the same procedures for all the
standards in sequence.
8.6
After all the standards are measured press ENTER to create the curve and the
screen return to the first with
* at “Name Curve”.
9.
Name Curve and Save Curve: Refer to “Name Curve and Save Curve” in “By
Factor” chapter in this manual.
10.
Display and Print Curve Parameters: Refer to “Dis & Print Curve” in By Known
Factors chapter in this manual.
UTILITY
In
MAIN MENU
, move * down to the left of
UTILITY
and press
ENTER
to go into
UTILITY
menu.