The instruments will automatically set to the proper wavelength as defined and blank the
reference(Press
0Abs/100%T
button to re-blank).
Goto preset wavelength
Blank automatically
Ready after Blanking
3. Insert the sample cuvette into the light-path (working position) and sample test result will
be displayed.
4.
Press PRINT or SAVE and follow the instruction to print or save the test results.
Note:
Run Test folder shows the pre-loaded tests ready for immediate test run. If the desired
test is defined and saved in “All Test List” but not yet loaded in “Run Test”, you need
to load the test method to “Run Test” before you can run the quantitative tests using
such method.
Load Test (method) to Run Test
1.
Select “All Test List at Quantitative menu.
2.
Selects the test method in “All Test List” to be loaded to Run Test.
3.
Press LOAD button to “copy and forward” the test method to Run Test.
4.
Exit from All Test List and enter to Run Test to check and make sure the selected test
method is listed.
Remove Test (method) from Run Test
1. At “Run Test” screen
scrolling * to the left of the test method you want to remove from the
Run Test folder.