Quick Start Guide UTS1000B/T Series
5
/
18
Instruments.uni-trend.com
System Information: access to the system menu and set up the relevant parameters
Reset (Default): press it to reset the setting to the default
Tracking Source (TG): the relevant setting of tracking source output terminal. Such as signal
amplitude, amplitude offset of tracking source. This key will light up when the trace source output
is working.
Single/Cont: press this key to perform single sweep. press it again to change it to continuous sweep
Touch/Lock: touch switch, press this key will indicate red light
5. Data Controller: direction key, rotary knob and numeric key, to adjust the parameter, such as center
frequency, start frequency, resolution bandwidth and make position
Note
Esc Key: If the instrument is in remote control mode, press this key to return to local mode.
6. Radio Frequency input terminal
(
RF input 50Ω
)
: this port is used to connect the external input signal, the
input impedance is 50Ω
(
N-Female connector
)
Warning
It is forbidden to load the input port with a signal that does not meet the rated value, and ensure that the
probe or other connected accessories are effectively grounded to avoid equipment damage or abnormal
function. RF IN port can only withstand an input signal power of no more than +30dBm or a DC voltage
input of 50V.
7. Tracking Source
(
TG SOURCE
)(
Gen Output 50Ω
)
: This N- Female connector is used to as the source
output of the built-in tracking generator. The input impedance is 50Ω.
Warning
It is forbidden to load input signals on the output port to avoid damage or abnormal function.
8. Loudspeaker: display analog demodulation signal and warning tone
9. Headphone Jack: 3.5 mm
10. USB Interface: to connect external USB, keyboard and mouse
11. ON/OFF Switch: short press to active the spectrum analyzer. In on-state, short press ON/OFF switch will
change the state to standby mode, all function will also be off.