Quick Start Guide                                                                                                                                                                                            UTS1000B/T Series 

 

16

 / 

18

 

Instruments.uni-trend.com 

 

Spectrum analyzer has 10.1 inch multipoint touch screen for various gesture operating, which includes,   

 

Tap the top right on the screen to enter the main menu. 

 

Slide up/down, left/right in waveform area to change the center frequency of X axis or reference level 
of Y axis. 

 

Zoom two points in waveform area to change the sweep width of X axis. 

 

Tap parameter or menu on the screen to select and edit it.   

 

Turn on and move the cursor. 

 

Use auxiliary quick key to perform common operation.   
Use [Touch/Lock] to turn on/off touch screen function. 

 

Remote Control 

The  UTS1000B/T  series  spectrum  analyzers  support  communication  with  computers  via  USB  and  LAN 
interfaces.  Through  these  interfaces,  users  can  combine  the  corresponding  programming  language  or 
NI-VISA,  using  the  SCPI  (Standard  Commands  for  Programmable  Instruments)  command  to  remotely 
program  and  control  the  instrument,  as  well  as  interoperate  with  other  programmable  instruments  that 
support the SCPI command set. 
For more information about the installation, remote control and programming, please refer to official site 
http://

 www.uni-trend.com

 UTS1000B/T Series Programming Manual. 

 

Help Information 

The spectrum analyzer's built-in help system provides help information for each function button and menu 
control key on the front panel. 

  Touch the left of the screen “

”, help dialog box will pop out on the center of the screen. Tap support 

function to get more detailed help description.   

  After help information displayed on the center of the screen, tap “×” or other key to close the dialog box. 

 

Troubleshooting 

 
This chapter lists the possible faults and troubleshooting methods of the spectrum analyzer. 
Please follow the corresponding steps to handle it, if these methods is not work, please contact UNI-T and 
provide your machine. 
Device information (acquisition method: [System] >System Information) 
 

1. After press the power soft switch, the spectrum analyzer still display a blank screen, and nothing is displayed.   

a.

 

Check whether the power connector is properly connected and the power switch is turned on.   

b.

 

Check whether the power supply meets the requirements. 

c.

 

Check whether the fuse of the machine is installed or blown. 

2. Press the power switch, if the spectrum analyzer still display blank screen and nothing is displayed. 

a.

 

Check the fan. If the fan rotates but the screen is off, the cable to the screen may be loose.   

Summary of Contents for UTS1000B Series

Page 1: ...Quick Start Guide UTS1000B T Series 1 18 Instruments uni trend com Quick Start Guide UTS1000B T Series Spectrum Analyzer...

Page 2: ...urchasing this brand new product In order to use this product safely and correctly please read this manual thoroughly especially the safety notes After reading this manual it is recommended to keep th...

Page 3: ...the defective products to the individual or entity that is declared in the guarantee In order obtain the warranty service customer must inform the defects within the applicable warranty period to UNI...

Page 4: ...e Trace set up trace line demodulation mode and trace line operation Marker this maker key is to select marked number type attribute tag function and list and to control the display of these markers P...

Page 5: ...uency input terminal RF input 50 this port is used to connect the external input signal the input impedance is 50 N Female connector Warning It is forbidden to load the input port with a signal that d...

Page 6: ...m hold The third line is display the detector type which includes S sampling detection P peak value N normal detection A average f trace operation All detection type are displayed in white letters Tap...

Page 7: ...orage File System user can use file system to save the correction limiting value measuring result screenshot trace state or other file into internal or external storage and it can be recall System inf...

Page 8: ...quency Internal Warning It is forbidden to load the input port with a signal that does not meet the rated value and ensure that the probe or other connected accessories are effectively grounded to avo...

Page 9: ...rface TCP IP port for remote control connecting 6 USB Device Interface Spectrum analyzer can use this interface to connect a PC which can be remote control by the software on the computer 7 Power Swit...

Page 10: ...roperty loss caused by the user s failure to follow the following safety precautions This device is designed for professional users and responsible organizations for measurement purposes Do not use th...

Page 11: ...switch is turned off this device is not completely disconnected from the AC power supply CAT I Secondary electrical circuit connected to wall sockets through transformers or similar equipment such as...

Page 12: ...ine voltage The line voltage of the line fuse of this device is correct Do not used to measure MAINS CIRCUIT Check all terminal rated values Please check all rated values and marking instructions on t...

Page 13: ...ted current 5A rated voltage 250V by the maintenance personnel authorized by UNI T Disassembly and cleaning There are no components available to operators inside Do not remove the protective cover Mai...

Page 14: ...on the rear panel and side panel of the instrument So please keep the air flowing through the vents of the instrument housing To prevent excessive dust from blocking the vents please clean the instru...

Page 15: ...standby mode 3 Press the switch on the front panel indicator lights up green and then the spectrum analyzer is powered on It takes about 30 seconds to initialize the boot and then the spectrum analyz...

Page 16: ...refer to official site http www uni trend com UTS1000B T Series Programming Manual Help Information The spectrum analyzer s built in help system provides help information for each function button and...

Page 17: ...t results and accuracy problems To achieve the performance listed in this manual you need a Check whether external device is properly connected and work b Have a certain understanding of the measured...

Page 18: ...or repair or acquire warranty form please contact the nearest UNI T sales and repair department In addition to permit provided by this summary or other applicable insurance guarantee UNI T does not pr...

Reviews: