User’s Manual Series DC Electronic Load
Instruments.uni-trend.com
26
/
33
8.
System (Menu)
The main menu divides into four parts: system setting, parameter setting, file operation and device
information. The parameter setting is only activated for the selected channel (such as needs to
use CH key to select the channel), other setting are all valid.
<Main Menu>
Press Shift and then Mode to enter system <Menu> setting interface, as shown in the Figure 8-1.
Figure 8-1 System Main Menu
8.1 System Setting
Move the cursor to
【
System Config
】
, press
【
Enter
】
key to enter < System Config > page, this
page includes the list setting as follows.
System Parameter
Setting
Description
Language
CHN/Englishh
Set the system language
Knob Active
ON/OFF
The data adjust by the rotary knob will take effect
immediately when the load is in activating state
Warning
ON/OFF
Turn on/off alarm sound
KeyVoice
ON/OFF
Turn on/off key sound
Initial Mode
Last/Default
Select Default, the electronic load enters CC test
interface
Select Last, the electronic load enters the test
mode at last time
Address
001~032
Set the communication address
Baud Rate
9600/19200/38
400/
57600/115200
Set the baud rate of RS232 port
Factory Set
Default
Restore to the factory setting