7.1.6
List Test
List test can switch different mode in rotation according to the set parameter. For power supply
products and charger device, etc., the multi-parameter mixed test can provide a more
comprehensive and in-depth understanding of the comprehensive working characteristics of the
tested products in practical applications.
Parameter of List Mode
Item
Parameter
Description
Group
1~60
Set the group number of list test file
for using
Step
1~16
Set the step for the group of list test
Repeat Times
0~99999
Set the repeat operating times for the
list file
Mode
Cont/Trig/ContEX/TrigEX
Set the switching mode and stop
mode for each step
Group Number
The internal Flash of electronic load can store 60 groups of list files. When set the list parameters, set
a group number at first. Press Shift and then the Save key to save the setting.
Operation Mode
Four operation mode: Cont/Trig/ContEX/TrigEX.
Cont Mode: The electronic load will stop the test if it over the limit or error occurs.
Trig Mode: The electronic will stop when a step is finished, and waits for trigger signal to continue next
step.
EX Mode: The electronic load will continue to next step even though it over the limit or error occurs.
Mode Parameter Setting
Item
Description
Mode CC/CV/CR/CP/Open/Short
Select the current step
Value
Set the constant value for
the mode
Set the constant value for the mode, default
constant value of Open/Short is 1
Time 300~999999ms
Set the load timing of each step, the range is
from 300 to 999999ms
Check OFF/Current/Voltage/Power Select the check item
Upper
The upper limit of check
item
Set the upper limit of check item
Lower
The lower limit of check
item
Set the lower limit of check item