User’s Manual Series DC Electronic Load
Instruments.uni-trend.com
20
/
33
trigger signal for each time and don’t need to set the timing of low value. Pulsed transient operation
as shown in the following figure.
Figure 7-1-5-2 Pulsed Transient Operation
3.
Reversal Mode
The electronic load will switch between the high and low values once for each trigger. At this time,
there is no need to set the timing for both high and low. And it will only switch to the other state
after each trigger is performed. Toggled transient operation as shown in the following figure.
Figure 7-1-5-3 Toggled Transient Operation
Operation steps
1.
After the instrument is boot-up, press CH key to select the channel, yellow wireframe area is the
selected channel and it will display the prompt of CH1 or CH2 at top of the screen (
single channel don’t need to choose.)
2.
Press Mode key to enter the selection interface, use rotary knob or direction key to select
dynamic mode and then press Enter key to enter test interface.
3.
Rotate the rotary knob to select the parameter to adjust, press Enter key to change and press it
again to save the setting.
4.
After the parameter is set, press ON key to enter dynamic test interface (press ON1 to enter CH1,
press ON2 to enter CH2).
5.
Press ON key to operating, press ON key again to stop the operation.