Service Manual UTG1000X Series
Instruments.uni-trend.com
14
/
20
2. Gently remove the front panel, as shown in the figure below.
Note: When the front panel is placed downward, it is necessary to avoid the knob cap to avoid damage
to the knob.
Removing the Front Panel Assembly
The following procedure describes the removal of the front panel.
Prerequisites:
To prevent electrostatic damage to components, wear a properly grounded antistatic wrist and foot
strap during installation, and use an antistatic mat in a tested antistatic environment.
Steps:
1
、
Place the cushion flat on the electrostatic table;
2
、
Place the instrument face down on a cushion to prevent damage to the screen and knobs;
3
、
Remove the connecting wire harness on the front panel; as shown in the figure below: