background image

 
                                                                                                   

 

 

 

7

into the LOCKED position as shown in Figure. Confirm that the sensor head is properly closed. 

 

 

 

 

Degaussing and Zero Setting 

When the key is pressed, the probe will demagnetize the core and set the output to zero voltage if 

it has been magnetized by switching the power on and off, or by an excessive input. Always carry out 
demagnetizing and Zero Setting before measurement and without current in the clamp. The 
demagnetizing and Zero Setting process takes about a few seconds. During demagnetizing and Zero 
Setting, a demagnetizing waveform is output. 

If degaussing and Zero Setting succeeds, the buzzer will make two short beeps. If degaussing and 

Zero Setting failed, the buzzer will make a single sound, for one second. 

Do not demagnetize while the conductor being measured is clamped. This could damage the 

components of the circuit being measured. Also, check that the conductor being measured is not 
clamped when supplying power to the current probe for the same reason. Demagnetized waveforms are 
generated when switching on the supply. 

 

 

Measurement 

 

Press the opening lever to open the sensor head. 

 

Align the sensor so that the probe’s current direction indication corresponds to the direction of 
current flow through the conductor to be measured. Also, align the clamp so that the conductor is 
in the center of the sensor aperture. 

 

Press the opening lever on the sensor head until the UNLOCK indication disappears. Check that 
the opening lever is firmly locked, and the sensor head securely closed. 

 

4.

 

  Safe Probing 

This device is designed to comply with Safety Standards and has been thoroughly tested for safety prior 
to shipment. However, mishandling during use could result in injury or death, as well as damage to the 
device. Be certain that you understand the instructions and precautions in the manual before use. We 
disclaim any responsibility for accidents or injuries not resulting directly from device defects. 
To avoid short circuits and potentially life-threatening hazards, follow these warnings and precautions: 

 

 

Never attach the clamp to a circuit that operates over the maximum rated voltage to earth. 

 

Please avoid clamping around bare conductors during measurement. 

 

While clamping and measuring, do not touch the clamp in front of the barrier or the 
conductor being measured. 

 

Be careful to avoid damaging the insulation surface while taking measurements. 

Summary of Contents for UT-P4000 Series

Page 1: ...www uni trend com DC AC Current Probes UT P4000 Series UT P4030D 30A DC 100MHz UT P4150 150A DC 12MHz UT P4500 500A DC 5MHz...

Page 2: ...ss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure p...

Page 3: ...istics 10 6 2 Mechanical characteristic 13 6 3 Environmental characteristic 14 7 Operating Method 14 7 1 Preparation before testing 15 7 2 Degaussing and Zero setting 15 7 3 Measuring method 15 8 The...

Page 4: ...capabilities Over current protection with dual indicators buzzer and LED High and low range selection Low current measurements Degaussing and automatic zero setting Digital set by panel soft push key...

Page 5: ...ul during measurement 2 Opening lever The operating lever used to open the sensor head Pull the lever to open the sensor jaw put in the cable under test and push the lever to lock the sensor head to m...

Page 6: ...Model Range Transfer ratio UT P4030D 30A 0 1V A 5A 1V A UT P4150 150A 0 01V A 30A 0 1V A UT P4500 500A 0 01V A 75A 0 1V A 10 Power supply socket External power supply socket standard with 12V 1A adap...

Page 7: ...the sensor head Align the sensor so that the probe s current direction indication corresponds to the direction of current flow through the conductor to be measured Also align the clamp so that the co...

Page 8: ...it from vibration or shock during transport and handling and be especially careful to avoid dropping Do not store or use the device where it could be exposed to direct sunlight high temperature humidi...

Page 9: ...ontinuous maximum continuous input for an extended period of time may result in activation of the safety circuit to prevent damage resulting from heating of the sensor At high ambient temperatures the...

Page 10: ...pears and check that the sensor head is properly closed If the sensor head is not properly closed an accurate measurement is not possible Accurate measurement may be impossible in locations subject to...

Page 11: ...0A 300Ap 500A 750A Current transfer ratio 5A 1V A 30A 0 1V A 75A 0 1V A 30A 0 1V A 150A 0 01V A 500A 0 01V A Measurable current 5A 1mA 30A 10mA 75A 10mA 30A 10mA 150A 100mA 500A 100mA Amplitude accura...

Page 12: ...e2 UT P4030D Figure3 UT P4030D Continuous maximum input measurement Input impedance VS Frequency curve Figure4 UT P4150 Amp Frequency curve Figure5 UT P4150 Figure6 UT P4150 Continuous maximum input m...

Page 13: ...put impedance VS Frequency curve 6 2 Mechanical characteristic Model UT P4030D UT P4150 UT P4500 Measurement conductor diameter max 5mm 20mm Cable length 1m 1 5m BNC Cable length 100cm Adapter dimensi...

Page 14: ...voltage might deviate because of the surrounding temperature so please be careful when testing sequentially The frequency of the current under test may cause resonance but this won t influence the tes...

Page 15: ...the high frequency current probe UT P4000 series adapter and oscilloscope Power up the UT P4000 probe and the green LED power indicator will be lighted Set the oscilloscope Ground the measuring mode...

Page 16: ...powered by standard adapter independent of oscilloscope power so it is very easy to use 9 2 Can UT P4000 series product measure small current A Yes For now the UT P4000 series current probe has two o...

Page 17: ...nt because the probe is not on the circuit yet The current under test should not surpass the limit value of the probe Please always maintain your probe and do not use it in the humid environment If th...

Reviews: