UNI-T UT-P4000 Series Manual Download Page 16

 
                                                                                                   

 

 

 

16

UT-P4030 is 0.1V/A (30A range), and then, when the voltage sensibility of the waveform monitor 

is 10mV/div, the current sensibility is 100mA/div. 

 

8.

 

The method to deal with abnormal situation 

Situation 

Possible reason 

Dealing method 

Can’t measure DC, or the value 

obtained is comparatively low in 

the frequency range 

Power is off 

Turn on power 

Oscilloscope set to AC coupling 

Set to DC coupling 

Sensor is not locked 

Please lock the sensor   

Auto degaussing or zero setting 

unsuccessful   

The probe is on the operating circuit under test 

when degaussing or zero setting is applied 

Turn off the circuit under 

test and zero set again. 

The amplitude is comparatively 

low in the frequency range 

The input resistance of the test equipment like 

oscilloscope is 50Ω 

Set the resistance over 

1MΩ 

 

 

9.

 

  Q&A 

9.1 Does UT-P4000 series fit the oscilloscope of any brand? 

A: UT-P4000 series has standard BNC interface can be applied to the oscilloscope of any brand. It is 

powered by standard adapter, independent of oscilloscope power, so it is very easy to use. 

9.2 Can UT-P4000 series product measure small current? 

A: Yes. For now, the UT-P4000 series current probe has two optional ranges, and one is for small 

current. The current resolution of the UT-P4030(D) is 1mA. When measuring small current, please 
accurately zero set and degaussing the probe, and do not change the position of the probe hand grip. To 
observe the waveform please set the bandwidth restriction of the oscilloscope to 20MHz to eliminate the 
interference of noise. When measuring extremely small current (a few mA for example), one could make 
a few more loop of cable around the probe and divide the result with number of loop to obtain the actual 
current value. 

9.3 Anymore tips? 

 

When measuring high frequency current, please do not let the current surpass the value shown by 

the curve of max peak current vs frequency. The max continuous current over the curve will burn 

the probe. 

 

To measure accurately, please degauss and zero set the probe, and make sure the probe is locked 

during the process. 

 

Set the input impedance of the oscilloscope to 1MΩ(default) 

Summary of Contents for UT-P4000 Series

Page 1: ...www uni trend com DC AC Current Probes UT P4000 Series UT P4030D 30A DC 100MHz UT P4150 150A DC 12MHz UT P4500 500A DC 5MHz...

Page 2: ...ss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure p...

Page 3: ...istics 10 6 2 Mechanical characteristic 13 6 3 Environmental characteristic 14 7 Operating Method 14 7 1 Preparation before testing 15 7 2 Degaussing and Zero setting 15 7 3 Measuring method 15 8 The...

Page 4: ...capabilities Over current protection with dual indicators buzzer and LED High and low range selection Low current measurements Degaussing and automatic zero setting Digital set by panel soft push key...

Page 5: ...ul during measurement 2 Opening lever The operating lever used to open the sensor head Pull the lever to open the sensor jaw put in the cable under test and push the lever to lock the sensor head to m...

Page 6: ...Model Range Transfer ratio UT P4030D 30A 0 1V A 5A 1V A UT P4150 150A 0 01V A 30A 0 1V A UT P4500 500A 0 01V A 75A 0 1V A 10 Power supply socket External power supply socket standard with 12V 1A adap...

Page 7: ...the sensor head Align the sensor so that the probe s current direction indication corresponds to the direction of current flow through the conductor to be measured Also align the clamp so that the co...

Page 8: ...it from vibration or shock during transport and handling and be especially careful to avoid dropping Do not store or use the device where it could be exposed to direct sunlight high temperature humidi...

Page 9: ...ontinuous maximum continuous input for an extended period of time may result in activation of the safety circuit to prevent damage resulting from heating of the sensor At high ambient temperatures the...

Page 10: ...pears and check that the sensor head is properly closed If the sensor head is not properly closed an accurate measurement is not possible Accurate measurement may be impossible in locations subject to...

Page 11: ...0A 300Ap 500A 750A Current transfer ratio 5A 1V A 30A 0 1V A 75A 0 1V A 30A 0 1V A 150A 0 01V A 500A 0 01V A Measurable current 5A 1mA 30A 10mA 75A 10mA 30A 10mA 150A 100mA 500A 100mA Amplitude accura...

Page 12: ...e2 UT P4030D Figure3 UT P4030D Continuous maximum input measurement Input impedance VS Frequency curve Figure4 UT P4150 Amp Frequency curve Figure5 UT P4150 Figure6 UT P4150 Continuous maximum input m...

Page 13: ...put impedance VS Frequency curve 6 2 Mechanical characteristic Model UT P4030D UT P4150 UT P4500 Measurement conductor diameter max 5mm 20mm Cable length 1m 1 5m BNC Cable length 100cm Adapter dimensi...

Page 14: ...voltage might deviate because of the surrounding temperature so please be careful when testing sequentially The frequency of the current under test may cause resonance but this won t influence the tes...

Page 15: ...the high frequency current probe UT P4000 series adapter and oscilloscope Power up the UT P4000 probe and the green LED power indicator will be lighted Set the oscilloscope Ground the measuring mode...

Page 16: ...powered by standard adapter independent of oscilloscope power so it is very easy to use 9 2 Can UT P4000 series product measure small current A Yes For now the UT P4000 series current probe has two o...

Page 17: ...nt because the probe is not on the circuit yet The current under test should not surpass the limit value of the probe Please always maintain your probe and do not use it in the humid environment If th...

Reviews: