background image

 

 

 

UPO1000CS Series

 

Digital Phosphor Oscilloscope 

Quick Start Guide 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REV 1 

2022.11.7 

 

 

Summary of Contents for UPO1000CS Series

Page 1: ...UPO1000CS SeriesDigital Phosphor Oscilloscope Quick Start Guide REV 1 2022 11 7...

Page 2: ...re protected by patent rights in China and foreign countries including issued and pending patents UNI T reserves the rights to any product specification and pricing changes UNI T reserves all rights L...

Page 3: ...oceed to the next step until you fully understand and meet the conditions stated in the Warning statement Caution Caution indicates the presence of a hazard It reminds users to pay attention to a cert...

Page 4: ...away from CAT IV circuit CAT III Primary circuit of large equipment directly connected to the distribution board and circuit between the distribution board and the socket three phase distributor circ...

Page 5: ...ly Please be sure to ground this product before it is powered on AC power supply Please use the AC power supply specified for this device Please use the power cord approved by your country and confirm...

Page 6: ...ersonnel of UNI T for testing Any maintenance adjustment or parts replacement must be done by the relevant personnel of UNI T Cooling Do not block the ventilation holes at the side and back of this de...

Page 7: ...r the first time 1 Check for Damages Caused by Transport If the packaging carton or the foam plastic cushions are severely damaged please contact the UNI T distributor of this product immediately 2 Ch...

Page 8: ...nter the normal interface 3 Connect Probe Use probe in the attachment and connect it BNC terminal to the channel 1 BNC terminal of the oscilloscope Connect the probe s main alligator clip to the Compe...

Page 9: ...yed waveforms Excessive Compensation Correct Compensation Insufficient Compensation Figure 2 Compensating Calibration of Probe If the displayed waveform is look like the above Insufficient Compensatio...

Page 10: ...r input terminal 2 Multipurpose knob 9 Horizontal control area 3 Function menu 10 Analog channel input terminal 4 Run Stop Automatic setting 11 Vertical control area 5 Trigger control 12 Control menu...

Page 11: ...11 14 5 Rear Panel Figure 4 Rear Panel Table 2 Description of Rear Panel NO Description NO Description 1 USB Device interface 3 Pass Fail output terminal 2 LAN interface 4 Safety lock 1 2 3 4...

Page 12: ...scale 10 LAN identification 3 Sampling rate Storage depth 11 USB host identification 4 Waveform identification 12 MATH state identification 5 Trigger position 13 CH2 state identification 6 Horizontal...

Page 13: ...trol modes a User defined programming b Use PC software instrument manager c Web Control 9 Troubleshooting 1 When press the soft power button the oscilloscope is blank screen a Check whether the power...

Page 14: ...ger level exceeds the waveform range b If it exceeds set the trigger level to the middle or set the mode to AUTO c Press AUTO button could automatically finish the above setting 6 Waveform refresh is...

Reviews: