a.

 

Check whether the mode at the trigger panel (TRIGGER) is on “Normal” or “Single” and whether the 
trigger level exceeds the waveform range.   

b.

 

If it exceeds

set the trigger level to the middle or set the trigger mode to Auto. 

c.

 

Press AUTO button could automatically finish the above setting. 

(6)

 

Waveform refresh is very slow: 
a.

 

Check whether the acquisition method is average and the average times are large. 

b.

 

Check whether the storage depth is the maximum. 

c.

 

Check whether the trigger release time is large. 

d.

 

Check whether the trigger is normal and the current time base is slow. 

e.

 

All the preceding causes slow refreshing of waveforms. Users can restore the factory settings so 
that the waveforms can be refreshed normally. 

11. 

  Appendix Contact Us 

If the use of this product has caused any inconvenience, if you in mainland China, you can contact UNI-T 
company directly. 
Service support: 8am to 5.30pm (UTC+8), Monday to Friday or via email. Our email address is infosh@uni-
trend.com.cn 
For product support outside mainland China, please contact your local UNI-T distributor or sales center. 
Many UNI-T products have the option of extending the warranty and calibration period, please contact 
your local UNI-T dealer or sales center. 

 

To  obtain  the  address  list  of  our  service  centers,  please  visit  our  official  website  http://www.uni-
trend.com 

Summary of Contents for MSO Series

Page 1: ...Quick Start Guide MSO UPO3000E Series Oscilloscope...

Page 2: ...221130 V1 00 Statement UNI T products are protected by patent rights in China and foreign countries including issued and pending patents UNI T reserves the rights to any product specification and pric...

Page 3: ...occur if the rules in the Warning statement are not properly executedorobserved Donotproceedtothenextstepuntilyoufullyunderstand and meet the conditions stated in the Warning statement Caution Caution...

Page 4: ...le tools home appliances etc Household appliances portable tools e g electric drill household sockets sockets more than 10 meters away from CAT III circuit or sockets more than 20 meters away from CAT...

Page 5: ...avoid electric shock the grounding conductor must be connected to the ground This product is grounded through the grounding conductor of the power supply Please be sure to ground this product before i...

Page 6: ...rsonal injury Caution Abnormality If this device may be faulty please contact the authorized maintenance personnel of UNI T for testing Any maintenance adjustment or parts replacement must be done by...

Page 7: ...Not support 3 Getting Started Guide This chapter introduces on using the oscilloscope for the first time the front and rear panels the user interface as well as the built in help system 3 1 General I...

Page 8: ...on the indicator of power soft button on the left corner on the front panel will be extinguished and no function If the power switch on the rear panel is turned on the indicator of power soft button o...

Page 9: ...the oscilloscope s compensating signal and ground terminal respectively Open CH1 and press the AUTO button View the displayed waveform as shown in Figure 2 Excessive Compensation Correct Compensation...

Page 10: ...3 Trigger control TRIGGER 3 Recording waveform 14 Horizontal control HORIZONTAL 4 Jog dial knob 15 Analog channel input port 5 Function menu 16 Vertical control VERTICAL 6 Numeric keypad 17 Menu contr...

Page 11: ...17 5 Rear Panel Figure 4 Rear Panel Table 2 Rear Panel No Description No Description 1 EXT Trig port 5 LAN 2 AUX Out and arbitrary waveform signal output 6 AC Power Input Socket 3 VIDEO Out 7 Power S...

Page 12: ...asily operating system with flexible and high sensitive touch screen features for great waveform display and excellent user experience Touch control function includes tap squeeze drag and drawing rect...

Page 13: ...e Gesture 7 3 Drag Use onefinger to pressanddrag the selected itemto the aimed positionas shownin Figure 9 Drag gesture can use for Drag waveform to change waveform displacement or offset Drag window...

Page 14: ...to one direction this gesture can quickly turn on off time measurement and voltage measurement cursor a Time measurement cursor Slide down in vertical to directly turn on time measurement cursor as s...

Page 15: ...e menu This symbol represents that multipurpose knob on the front panel can be used to select and adjust parameter This symbol represents that there are several options This symbol represents that the...

Page 16: ...steps restart the oscilloscope e If the instrument still cannot work please contact UNI T 2 After signal acquisition signal of waveform does not appear on the screen a Check whether BNC port is conne...

Page 17: ...trigger is normal and the current time base is slow e All the preceding causes slow refreshing of waveforms Users can restore the factory settings so that the waveforms can be refreshed normally 11 Ap...

Reviews: