USER MANUAL - URS600CE AND URS600CESS JULY 2019
30
Fault code Flashing LED light
Description of the fault Required Action
24
Ready L
2x
Ready H
4x
Code 23 lasts more
than 10 minutes.
The heating circuit is automatically
turned off but the fan will continue to
cool until the FAN light goes out.
Once the URS600CE has cooled
down, disconnect the power cord in
a safe manner.
Remove the remaining solution/
residue from the distillation tank.
1. Inspect and clean the condenser
on the outside, so that all the cooling
fins are clean and free (see section
4.3).
2. Check that the URS600CE has
at least 15 cm free space around it
and the temperature in the room is
between 5 and 40 °C.
3. Check that the fan motor is
working properly. To do this,
reconnect the URS600CE to the
power supply in the empty state and
switch the URS600CE on briefly. The
fan motor should switch on.
If these steps have been completed
without problems, run a test cycle
with 15 litres of clean water to test
whether the problem has been fixed.
If this doesn’t work, contact the
service/repair department of Uni-ram
Corporation
31
3x
1x
The solvent does not
reach the boiling point
within 45 minutes.
Wait until the FAN light goes out and
the URS600CE has cooled down.
Empty and clean the distillation tank,
replace the liner bag and fill the tank
with at least 6 litres of solvent.
The set temperature is
too low for the type of
solvent.
Wait until the FAN light goes out and
the URS600CE has cooled down.
Check the boiling temperature of
your solvent. Set the URS600CE to a
higher temperature if this is allowed.
Boiling temperature of
the solvent is too high.
You cannot recycle this solvent.
32
3x
2x
Reset occurred due to
a power failure.
Press the OFF and then the ON
button to continue the process.
Normally, the URS600CE is not
faulty.