PHENOMENON
TEST METHOD IMMUNITY TEST LEVEL
Electrical fast transient/
burst immunity
- AC power ports
EN 61000-4-4
+/- 2 kV
+100 kHz repetition frequency
Electrical fast transient/
burst immunity
- DC power ports
EN 61000-4-4
Not applicable
Electrical fast transient/
burst immunity
- I/O SIO/SOP ports
EN 61000-4-4
Not applicable
Surge immunity
- AC power ports
EN 61000-4-5
Line-to-line: +/- 0.5, 1 kV line to line
Line-to-ground: Not applicable, the
system is a Class II device
Surge immunity
- DC power ports
EN 61000-4-5
Not applicable
Surge immunity
- I/O SIO/SOP ports
EN 61000-4-5
Not applicable
Immunity to conducted
disturbances induced by RF
fields - AC power ports
EN 61000-4-6
3 V (6 V in ISM bands and amateur radio
bands
a
)
0.15-80 MHz
80% AM 2 Hz
Immunity to conducted
disturbances induced by
RF fields - DC power ports
EN 61000-4-6
Not applicable