Immunity
PHENOMENON
TEST METHOD IMMUNITY TEST LEVEL
Electrostatic discharge
immunity
EN 61000-4-2
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
Electrostatic discharge
immunity - patient
coupling ports
EN 61000-4-2
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
Electrostatic discharge
immunity
- I/O SIO/SOP ports
EN 61000-4-2
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
Radiated RF
electromagnetic
field immunity
EN 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM 2 Hz
Immunity to proximity
fields from RF wireless
communication
equipment
EN 61000-4-3
385 MHz, 27 V/m, 18 Hz PM (50% duty cycle
square wave)
·
450 Mhz, 28 V/m, FM +/- 5 kHz
dev.,1 kHz sine
·
710 MHz, 9 V/m, 217 Hz PM
(50% duty cycle square wave)
·
745 MHz, 9
V/m, 217 Hz PM (50% duty cycle square wave)
·
780 MHz, 9 V/m, 217 Hz PM (50% duty cycle
square wave)
·
810 MHz, 28 V/m, 18 Hz PM
(50% duty cycle square wave)
·
870 MHz, 28
V/m, 18 Hz PM (50% duty cycle square wave)
·
930 MHz, 28 V/m, 18 Hz PM (50% duty cycle
square wave)
·
1720 MHz, 28 V/m, 217 Hz PM
(50% duty cycle square wave)
·
1845 MHz, 28
V/m, 217 Hz PM (50% duty cycle square wave)
·
1970 MHz, 28 V/m, 217 Hz PM (50% duty cycle
square wave)
·
2450 MHz, 28 V/m, 217 Hz PM
(50% duty cycle square wave)
·
5240 MHz, 9
V/m, 217 Hz PM (50% duty cycle square wave)
·
5500 MHz, 9 V/m, 217 Hz PM (50% duty cycle
square wave)
·
5785 MHz, 9 V/m, 217 Hz PM
(50% duty cycle square wave).