24/7 EEG™ SubQ
USER MANUAL | ENGLISH |
43
PHENOMENON
TEST METHOD IMMUNITY TEST LEVEL
Immunity to conducted
disturbances induced
by RF fields - I/O SIO/SOP
ports
EN 61000-4-6
Not applicable
Immunity to conducted
disturbances induced
by RF fields - Patient coupling
ports
EN 61000-4-6
Not applicable
Power frequency
magnetic field
immunity
EN 61000-4-8
30 A/m
50 Hz
Voltage dips, short
interruptions and
voltage variations
immunity
EN 61000-4-11
0% UT: 0.5 cycle at 0
o
, 90
o
, 135
o
, 180
o
, 225
o
,
270
o
and 315
o
0% UT: 1 cycle at 0
o
70% UT: 10 cycle at 0
o
0% UT: 25 cycle at 0
o
Electrical transient
conduction along
supply lines
ISO 7637-2
Not applicable
a
The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.5 67 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz. 3.5 MHz to 4.0
MHz. 5.3 MHZ to 5.4 MHz. 7 MHZ to 7.3 MHz. 10.1 MHz to 10.15 MHz. 14 MHz to 14.2 MHz. 18.07 MHz
to 18.17 MHz. 21.0 MHZ to 21.4 MHz. 24.89 MHz to 24.99 MHz. 28.0 MHz to 29.7 MHz and 50.0 MHz
to 54.0 MHz.