6
1.6.6
Leakage of hot cooling water
If you inadvertently operate this pump without flowing water, stop the pump
immediately and keep away from it.
If you inadvertently operate this pump without flowing water, stop the pump
immediately and keep away from it.
Hot steam may blast out of the pump's cooling water port.
Install an interlock on the cooling water pathway.
If you operate this pump without flowing cooling water, hot steam may blast out of
the pump's cooling water port.
Wait until the pump temperature drops.
After stopping this pump and confirming that the pump temperature has dropped,
remove and inspect the pump.
1.7
Safety Data Sheet
Carefully read the safety data sheet.
Obtain and carefully read the safety data sheet (SDS) before using this pump.
If lubricant adheres to the skin or enters the eyes, follow the first aid procedure
described on the
SDS.
Use the specified lubricant.
Use of a non-specified lubricant will affect the pump's performance and service life
as well as void the pump's warranty coverage.
Do not use a chemical substance (lubricant) not specified in this document.
Pump fluid
ULVOIL
B-6
(ULVAC)
The SDS describes the chemical substances that may be used or touched to operate this pump. To
understand the characteristics of hazards, carefully read the SDS. Obtain the latest version of the SDS
from our nearest sales office or dealer.
The SDS provides reference information to ensure safe handling of dangerous or hazardous chemical
substances.
Anyone handling lubricant must always obtain the latest SDS and understand that measures suitable
for the actual handling and other situations in question must be taken under their own responsibility
by referring to the SDS before using the lubricant. The SDS itself does not guarantee safety