5.2.
I/O output signal
The sensor error and setpoint signals are output from the I/O connector of this unit
in open collector format.
Photocoupler rating [30V
MAX
、
50mA
MAX
、
70mW]
The internal circuit is as follows.
Diagram of output signal internal circuit
Sensor error signal
The sensor error signal is a function that outputs as a signal that the filament of
the sensor is disconnected.
If a sensor error occurs, the signal will be a Lo output.
I/O connector: 2pin [sensor error] - 15pin [GND]
If a sensor error occurs, the POWER/ERROR LED will light red and the pressure signal
output will be 9.0V or higher.
Note
Precautions when detecting disconnection
If the main component of the measuring gas is a gas with high
thermal conductivity such as hydrogen (H
2
) or helium (He) gas,
and the measurement is about 10,000 Pa or more, it may be judged
that the filament is disconnected even if the filament is not
disconnected.
Setpoint operation signal
Setpoint is a function to output a signal to the outside or turn on the LED when
the pressure drops below a certain set pressure. The set pressure value is called
"setpoint". If the measured pressure value drops below the setpoint, this signal
will be a Lo output.
For the setting method of the setpoint, refer to Chapter 6 of this manual and
Chapter 11 of this manual.
+5V
+24V
A
1
2
A
1
2
A
1
2
2:ERROR signal ON/OFF
3:SETPOINT signal-1 ON/OFF
15:GND
11:SETPOINT signal-2 ON/OFF
Internal circuit
PLC or PC
Internal circuit
Summary of Contents for G-TRAN SW100-A
Page 7: ...SK00 9663 EI 003 00 VII Revision history Date Number Reason 2022 10 3 00 First edition ...
Page 69: ...SK00 9663 EI 003 00 69 15 EC DECLARATION OF CONFORMITY ...
Page 70: ...SK00 9663 EI 003 00 70 16 UK DECLARATION OF CONFORMITY ...
Page 71: ...SK00 9663 EI 003 00 71 17 Certificate of decontamination ...
Page 72: ...SK00 9663 EI 003 00 72 18 Drawings 18 1 SW100 A R dimensions ...
Page 74: ...SK00 9663 EI 003 00 74 18 3 Display unit cable L ...