SK00-9663-EI-003-00
30
7.4.
Command
Reading measurement values and status
Command :
AD0
AD1
D
CHKH
CHKL
CR
Returning format from this unit to PC
:
AD0
AD1
D
X . X X E ± X X
SH
SL
CHKH
CHKL
CR
The part of
「
X
.
XXE±XX
」
is the measured pressure value.
Example 1
)3.00E+03 ⇒ 3.00×10
+3
Pa
Example 2
)5.00E+00 ⇒ 5.00×10
+0
Pa
Example 3
)
4.00E-
01 ⇒ 4.00×10
-1
Pa
When
「
E
.
EEE
+
EE
」
is returned, the filament is disconnected,
When
「
F
.
FFE
+
FF
」
is returned, the pressure is above the measurable pressure
range.
For the status “SH” and “SL”, refer to Section 7.3 of this manual.
Zero point adjustment command
Command :
AD0
AD1
ZER
CHKH
CHKL
CR
Adjustable pressure range is about ± 1Pa.
If the pressure cannot be adjusted, ‘n’ will be returned.
After adjustment, read the measured value and check it.
CPU processing time is required. Please send the following command after
1.5seconds or more have passed.
Atmospheric pressure adjustment command
Command :
AD0
AD1
ATM
CHKH
CHKL
CR
Adjustable pressure range is about 1×10
+4
Pa to 2×10
+5
Pa
If the pressure cannot be adjusted, ‘n’ will be returned.
After adjustment, read the measured value and check it.
CPU processing time is required. Please send the following command after
1.5seconds or more have passed.
Zero point, atmospheric pressure adjustment clear command
Command :
AD0
AD1
CLR
CHKH
CHKL
CR
If the command is received normally, the normal receiving ‘o’ will be
returned.
CPU processing time is required. Please send the following command after
1.5 seconds or more have passed.
Reading status
Command :
AD0
AD1
SR
CHKH
CHKL
CR
Returning format from this unit to PC
:
AD0
AD1
S
SH
SL
CHKH
CHKL
CR
For the status “SH” and “SL”, refer to Section 7.3 of this manual.
Summary of Contents for G-TRAN SW100-A
Page 7: ...SK00 9663 EI 003 00 VII Revision history Date Number Reason 2022 10 3 00 First edition ...
Page 69: ...SK00 9663 EI 003 00 69 15 EC DECLARATION OF CONFORMITY ...
Page 70: ...SK00 9663 EI 003 00 70 16 UK DECLARATION OF CONFORMITY ...
Page 71: ...SK00 9663 EI 003 00 71 17 Certificate of decontamination ...
Page 72: ...SK00 9663 EI 003 00 72 18 Drawings 18 1 SW100 A R dimensions ...
Page 74: ...SK00 9663 EI 003 00 74 18 3 Display unit cable L ...