SK00-9663-EI-003-00
26
6.
Setpoint settings
This chapter describes the setting of the set point.
Setpoint is a function to output a signal to the outside or turn on the LED when
the pressure drops below a certain set pressure. The set pressure value is called
"set point".
The setpoint is set by connecting a Windows/Android device with UL-MOBI installed
to SW100 with a USB cable and using UL-MOBI.
The setpoint of SW100-A can be set at any time, but careless setting change may
lead to failure of the vacuum device, so change when 24V is energized requires
password input.
The setpoint of SW100-R can be set only when SW100-R is not energized at 24V.
For the UL-MOBI download method and setpoint setting method, refer to this chapter and
Chapter11 of this manual.
In addition, it is possible to two setpoints, setpoint1 and setpoint2.
※
The setpoint setting is invalid for 4seconds after starting SW100.
6.1.
Setpoint ON/OFF pressure
The set point has a hysteresis in the pressure to turn on and the pressure to turn off.
Pressure value to turn on : Set value
Pressure value to turn off: Set value + 10%
Setpoint hysteresis