Problem
Possible Cause
Corrective Action
(10) AUTO READY button
blinks.
The level sensor cable of the
target dewar is disconnected
or not connected.
Check the level sensor cable
of the target device.
The level sensor of the target
device breaks down.
Please contact us.
(11) MANU START button
blinks.
Only 30 seconds left until the
time limit of 10 minutes of
manual dispense.
Press MANU START button
again to continue dispending
manually.
(12) AUTO READY button
and AUTO SUPPLY
button blink
alternately.
The time limit of automatic
dispense is over, but the level
sensor of the target device
send the signal to supply.
Press AUTO READY button
to cancel automatic transfer
mode and check the liquid
nitrogen level in the target
device. Please contact us if
any problem is found.
(13) Other failures
Please contact us.