-17-
Turning Direction
Turning Direction plate
Cap
Inlet
Outlet
3.2 Test operation
CAUTION
Danger of cap to fly
Remove cap from inlet and outlet.
Operation with cap being fitted can cause cap to fly by intake or exhaust impetus,
resulting in accident, failure, or bodily injury from contact with flying objects.
Remove cap
Danger of foreign matter entering inlet
When checking turning direction, be careful not to enter foreign matter into an
inlet.
Foreign matter entering inlet can cause failure.
Prevent foreign
matter from entering
Danger of overheating
Check that cooling fan is turning and cooling air is flowing.
If not, it can cause accident, failure or bodily injury such as burns due to overheating.
Check fan
Test operation
①
Open inlet and outlet
Remove caps (2 places)
from inlet and outlet of
vacuum pump.
②
Check turning direction
Open inlet, turn on electrical
source to start operating
vacuum pump, and Check
that air comes out from outlet.
(Vacuum pump turns
counter-clockwise when
viewed from motor side.)
If air does not come out from
outlet, vacuum pump may turn
in reverse.
In that case, stop vacuum pump,
turn off main electrical source and change 2 out of 3 cords of electric source
connection and change turning direction to correct one.
If you fit pump to vacuum system and control operation of vacuum pump by
remote control,
first check pump itself for turning direction
and then fit it to
vacuum system.