QSx Operating Manual
28
C6054101 (June 2020)
The screen displays:
“Unit Overheat”
“Please Wait”
“Switch Off to”
“reset”
“menu”
The QSx has detected that the ultrasonic
generator on the unit has operated above the
rated temperature and has disabled the
ultrasonic function. This also disables a cycle
from starting. The most likely cause is the
liquid evaporating when the level sensor is
switched off. Switch the unit off, allow the unit
to cool for a few minutes, top up the liquid
above the level sensor and switch the unit
back on.
The screen displays:
Over Temp
The QSxhas detected that the water in the
bath is above 75ºC and has aborted the cycle.
Please allow the liquid to cool or drain some
fluid and replace with cooler liquid.
If your problem persists, it may be possible to rectify any issues over the
telephone. Our dedicated service and product support personnel may be
able to troubleshoot your problem remotely, thus causing minimal
disruption.
Contact our After Sales Department:
+44 (0) 845 330 4238
+44 (0) 2920 83 7336
Please have to hand your model and serial number together with
information on the problem prior to contacting us.
If we are unable to solve your problem over the phone, we may suggest
returning your product to the distributor where purchased or directly to
Ultrawave. Where appropriate, we operate a Return To Base (RTB)
warranty and repairs policy.